Minimum cycle times for steam sterilization cycles


Type of Sterilizer

Item

 

Exposure time at 250°F (121°C)

 

Exposure time at 270°F (132°C)

 

Drying time

 

 

 

 

 

 

 

 

 

Gravity displacement

Wrapped instruments

 

30 min

 

15 min

 

15-30 min

 

 

 

 

 

 

 

 

Gravity displacement

Textile packs

 

30 min

 

25 min

 

15 min

Gravity displacement

Wrapped utensils

 

30 min

 

15 min

 

15-30 min

Dynamic-air-removal (e.g., prevacuum)

Wrapped instruments

 

Not Applicable

 

4 min

 

20-30 min

Dynamic-air-removal (e.g., prevacuum)

Textile packs

 

Not Applicable

 

4 min

 

5-20 min

Dynamic-air-removal (e.g., prevacuum)

Wrapped utensils

 

Not Applicable

 

4 min

 

20 min



Flash steam sterilization parameters 

 

Type of Sterilizer

Load configuration

Temperature

 

Time

 

Gravity displacement

 

Nonporous items only (i.e., routine metal instruments, no lumens)

132°C (270°F)

3 minutes

Gravity displacement

 

Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together

132°C (270°F)

10 minutes

Dynamic-air-removal (e.g., prevacuum)

 

 

Nonporous items only (i.e., routine metal instruments, no lumens)

132°C (270°F)

3 minutes

Dynamic-air-removal (e.g., prevacuum)

 

 

Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together

132°C (270°F)

4 minutes

Steam-flush pressure-pulse

 

Nonporous or mixed nonporous/porous items

132°C (270°F) Manufacturers’ instruction

4 minutes